CSCA Chemistry · Experiments & Applications · Section 1 of 14
Laboratory Apparatus and Their Uses
Why apparatus matters
Many experiment questions are really questions about apparatus: which vessel can be heated, which instrument is precise enough, which must be checked for leaks, which must be rinsed with its own solution. Learn each piece by its job and by its rules.
Vessels for holding and heating
- Test tube (试管): small-scale reactions; heated directly, no more than 1/3 full of liquid.
- Beaker (烧杯): dissolving solids, mixing, heating larger volumes; heated on a gauze (陶土网, formerly 石棉网).
- Conical flask (锥形瓶): the titration vessel, because its narrow neck lets you swirl without splashing; also a reaction vessel in gas generators. Heated on a gauze.
- Round-bottom or distillation flask (圆底烧瓶 / 蒸馏烧瓶): heating liquids for reactions and distillation; heated on a gauze and filled between 1/3 and 2/3.
- Evaporating dish (蒸发皿): evaporating a solution to leave a solid; heated directly and stirred with a glass rod.
- Crucible (坩埚): strongly heating solids, for example driving the water out of CuSO₄·5H₂O. It sits on a pipe-clay triangle (泥三角) on a tripod, is moved with crucible tongs (坩埚钳) and cools in a desiccator (干燥器) so that the solid cannot take up moisture again. A porcelain crucible is not used to melt NaOH, which attacks the SiO₂ in porcelain; an iron crucible is used.
| Heated directly | Heated on a gauze | Never heated |
|---|---|---|
| test tube, evaporating dish, crucible, combustion spoon | beaker, flask, conical flask | measuring cylinder, volumetric flask, burette, reagent bottle |
Instruments for measuring
- Pan balance (托盘天平): reads to 0.1 g. Object on the left pan, weights on the right (左物右码): mass of object = weights + rider (游码) reading. If the two are swapped, the object's mass is weights − rider, so 5.4 g "weighed" with 5 g of weights and the rider at 0.4 g is really only 4.6 g. Corrosive or water-absorbing solids such as NaOH are weighed in a small beaker, not on paper.
- Measuring cylinder (量筒): reads to 0.1 mL; the scale increases upward and has no zero mark. Choose the smallest cylinder that holds the volume in one go (9.5 mL → a 10 mL cylinder). Never dissolve, dilute, react or heat anything in it.
- Burette (滴定管): reads to 0.01 mL; the zero is at the TOP and the numbers increase downward. The acid burette (酸式, glass stopcock) holds acids and oxidising solutions such as KMnO₄, which would attack rubber. The base burette (碱式, rubber tube with a glass bead) holds alkalis, which would make a glass stopcock stick. Burettes with a PTFE stopcock can hold either.
- Pipette (移液管): delivers one fixed volume, such as 25.00 mL, accurately.
- Volumetric flask (容量瓶): makes a solution up to one exact volume. It has a single ring mark, a fixed size (commonly 50, 100, 250, 500 or 1000 mL) and a stopper. It is never used to dissolve, react, heat or store anything.
Apparatus for separation
- Funnel (漏斗) with filter paper: filtration.
- Separating funnel (分液漏斗): separating two immiscible liquids, and adding a liquid to a reaction at a controlled rate.
- Condenser (冷凝管): turns vapour back into liquid. Cooling water enters at the lower end and leaves at the upper end. A straight condenser is used for distillation; a bulb or coil condenser, held upright, for reflux.
- Long-stem funnel (长颈漏斗): adds liquid to a gas generator; its tip must dip below the liquid, or the gas escapes through it.
Two checklists
Check for leaks before use (检漏): the volumetric flask, the separating funnel and the burette, that is, everything with a stopper or tap that must hold liquid.
Rinse with the solution it will hold (润洗): the burette and the pipette only. The conical flask is NOT rinsed with the sample and the volumetric flask is NOT rinsed with the solution, because the extra solute left on the walls would cause an error. A conical flask or volumetric flask that is wet with distilled water causes no error.
Key takeaway: Know each piece by its job and its rules: test tubes, evaporating dishes and crucibles are heated directly, beakers and flasks on a gauze, measuring glassware never. Precision is 0.1 g (pan balance), 0.1 mL (measuring cylinder) and 0.01 mL (burette, pipette). Leak-test anything with a stopper or tap, and rinse only the burette and pipette with their own solution.
Key Formulas
Precision: pan balance 0.1 g | measuring cylinder 0.1 mL | burette, pipette 0.01 mL | Leak-test: volumetric flask, separating funnel, burette | Rinse with own solution: burette, pipette
Worked Example 1
Example 1: A student records the volume of NaOH used in a titration as 22.6 mL. What is wrong?
Step 1: A burette is read to 0.01 mL, so the record must have two decimal places.
Step 2: It should be written 22.60 mL. A value with one decimal place, such as 22.6 mL, is the precision of a measuring cylinder.
Worked Example 2
Example 2: Which burette should hold acidified KMnO₄ solution, and which should hold NaOH solution?
Step 1: KMnO₄ is a strong oxidising agent and attacks the rubber tube of a base burette, so it goes in the acid burette (glass stopcock).
Step 2: NaOH attacks glass and makes a ground-glass stopcock stick, so it goes in the base burette.
Worked Example 3
Example 3: Choose the apparatus to heat 2 g of CuSO₄·5H₂O strongly until all its water is driven off, and to weigh the residue accurately.
Step 1: Strong heating of a solid needs a crucible on a pipe-clay triangle on a tripod, heated directly.
Step 2: Move the hot crucible with crucible tongs and let it cool in a desiccator, so that the anhydrous CuSO₄ cannot take up water from the air.
Step 3: Weigh, heat again, cool and weigh again until two masses agree (heating to constant mass).
Exam Tips
When an option says a piece of glassware is used to dissolve, dilute, react or heat something, check that it is not a measuring cylinder or a volumetric flask: those two only measure.
Read the decimal places in the question: 25.00 mL points to a burette or pipette, 25.0 mL to a measuring cylinder, 2.0 g to a pan balance.
Common Mistakes
Rinsing the conical flask with the sample. Only the burette and pipette are rinsed with their own solution; a rinsed conical flask holds extra sample and gives a high result.
Putting NaOH in an acid burette, or KMnO₄ in a base burette. Alkali makes a glass stopcock stick; an oxidant attacks rubber.
Heating a beaker directly in the flame. Beakers and flasks need a gauze to spread the heat.
Try This Yourself
Which of these can be heated directly in a flame: beaker, crucible, volumetric flask, test tube, conical flask?
Show the answer
The crucible and the test tube. The beaker and the conical flask need a gauze, and the volumetric flask is never heated.